QUANTIFICATION OF TOTAL MECLIZINE HYDROCHLORIDE CONTENT IN PHARMACEUTICAL FORMULATION BY UV SPECTROPHOTOMETRIC AIDED MULTIVARIATE CALIBRATION TECHNIQUE Authors: Nithyasri K , KAVITHA J* AND MANIKANDAN K
ABSTRACT
Present abstract employs the use of Multivariate Calibration Technique for the assessment
of Meclizine hydrochloride in the marketed pharmaceutical formulation. The multivariate
calibration method was developed by utilizing a linear regression equation to analyze the
relationship at five distinct concentration levels and recording its absorbance at five different
wavelengths close to the drug's maximum absorbance. Over a concentration range of 5–15
g/mL, the technique was shown to be linear with a correlation coefficient value greater than
0.998. Meclizine hydrochloride's ?max was discovered to be 230 nm. According to ICH
protocol Q2 (R1), the created approach was determined to be easy, quick, accurate, and
exact. By removing the discrepancies the resulting from instrumental or experimental
conditions, this statistical approach produces optimal results.
Keywords: Meclizine hydrochloride, Antihistamine, Multivariate calibration
technique, ICH protocol, Validation, UV Spectrophotometry Publication date: 01/10/2024 https://ijbpas.com/pdf/2024/October/MS_IJBPAS_2024_8379.pdfDownload PDFhttps://doi.org/10.31032/IJBPAS/2024/13.10.8379