SIMULTANEOUS ESTIMATION OF MOMETASONE FUROATE AND FLUTICASONE PROPIONATE BY USING UV SPECTROSCOPY Authors: Jadhav MS , TALELE GS AND MALPURE PS*
ABSTRACT
A simple, specific, accurate and precise UV spectrophotometric method has been developed
for the simultaneous estimation of Mometasone Furoate monohydrate and Fluticasone
Propionate in pharmaceutical dosage form. The absorption maxima of the Mometasone
Furoate monohydrate and Fluticasone propionate were found to be 248 nm and 236 nm
respectively using Acetonitrile as solvent. This method obeys Beer’s law in the employed
concentration range of 2-10?g/ml and 1-6?g/ml for Mometasone Furoate monohydrate and
Fluticasone propionate respectively. Different analytical performance parameters such as
linearity, precision, accuracy, limit of detection (LOD) and limit of quantification (LOQ)
were determined according to ICH guidelines. The LOD of Mometasone Furoate
monohydrate and Fluticasone propionate was found to be 0.060?g/ml and 0.42?g/ml
respectively and LOQ of Mometasone Furoate monohydrate and Fluticasone propionate was
found to be 0.182?g/ml and 0.321?g/ml respectively.
Keywords: UV, Simultaneous estimation, Mometasone Furoate, Fluticasone Propionate,
linearity Publication date: 01/12/2023 https://ijbpas.com/pdf/2023/December/MS_IJBPAS_2023_7162.pdfDownload PDFhttps://doi.org/10.31032/IJBPAS/2023/12.12.7162