COMPARATIVE STUDY OF ULTRAVIOLET SPECTROSCOPY AND HIGH PERFORMANCE THIN LAYER CHROMATOGRAPHIC METHODS FOR ELUXADOLINE Authors: Damle MC* And Pardeshi RR
ABSTRACT
The aim of present research work is to develop and validate UV-spectrophotometric and
High Performance Thin Layer Chromatography (HPTLC) methods for Eluxadoline. The
estimation of Eluxadoline was performed by both UV and HPTLC method with detection
wavelength of 238 nm. Chromatographic separation was carried on TLC aluminium plates
pre-coated with silica gel G 60 F254 with methanol: ethyl acetate: triethylamine (5:5:0.1 v/v)
as mobile phase. The retardation factor (RF) was observed to be 0.45±0.02.The developed
methods were successfully validated as per ICH Q2 (R1) guideline. Methods were found to
be linear within the range of 2-12µg/ml with correlation coefficient R2= 0.9963 (for UV
method) and 200-1000 ng/band with correlation coefficient R2 = 0.9982 (for HPTLC
method). The methods were precise as %RSD < 2 % and accurate as % recovery were
found to be in range of 98-102 %. LOQ of developed methods were found to be 2.27 µg/ml
and 87.98 ng/band for UV and HPTLC method respectively. The results of assay and
accuracy studies of both methods were evaluated by student t test. Statistical analysis shows
that there is no significant difference (p>0.05) between UV and HPTLC methods regarding
accuracy and assay results.
Keyword: Eluxadoline, HPTLC, Student t test, UV method Publication date: 01/11/2022 https://ijbpas.com/pdf/2022/November/MS_IJBPAS_2022_6618.pdfDownload PDFhttps://doi.org/10.31032/IJBPAS/2022/11.11.6618