REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC METHOD DEVELOPMENT AND VALIDATION FOR ESTIMATION OF APREMILAST IN PHARMACEUTICAL FORMULATION Authors: Srinivas Ampati And Hareesh Dara
ABSTRACT
In this current investigation effort, an effort was made to build up an uncomplicated,
speedy, truthful as well as robust HPLC technique for the assessment of Apremilast in its
tablet dosage form. Reverse phase high performance liquid chromatographic analysis was
carried out on isocratic system. The column used for the investigation was Phenomenex
C18 (250 mm× 4.6mm, 5µm) with ambient temperature. The optimized mobile phase was
Methanol: Acetonitrile in the ratio of (20:80 %V/V). The detection was carried out at a
wavelength of 230nm using a flow rate of 1ml/min. The urbanized technique was
validated for validation constraints like linearity, specificity, accuracy, precision as per
ICH strategy. The %RSD for all constraints was well within the limits, which indicates
the validity of the technique in addition to the assay results obtained are in reasonable
conformity with the label claim of the marketed formulation. Thus, the conventional
scheme can be anticipated for repetitive investigation of this drug in laboratories and for
superiority purposes.
Keywords: Apremilast, Acetonitrile, HPLC, Linearity Publication date: 25/09/2021 https://ijbpas.com/pdf/2021/September/MS_IJBPAS_2021_SEPT_SPCL_1041.pdfDownload PDFhttps://doi.org/10.31032/IJBPAS/2021/10.9.1041